Kabinet ujian suhu 600 liter 5 K/minVötsch
VTS 7060-5 -70°C bis + 180°C
Kabinet ujian suhu 600 liter 5 K/min
Vötsch
VTS 7060-5 -70°C bis + 180°C
Keadaan
Terpakai
Lokasi
Borken 

Gambar menunjukkan
Tunjukkan peta
Data mesin
Harga & Lokasi
- Lokasi:
- Einsteinstraße 8a, 46325 Borken, Deutschland

Hubungi
Butiran tawaran
- ID iklan:
- A20544940
- Nombor rujukan:
- 24880
- Kemas kini:
- terakhir pada 17.11.2025
Penerangan
Vötsch
Type: VTS 7060-5
Halogen Bulb 24V, 50W
Vötsch temperature test chamber with color touch panel and S!MPAT
Temperature range: -70°C to +180°C, water-cooled
Temperature deviation over time: ±0.1 to ±0.5 K
Temperature deviation spatially: ±0.5 to ±2.0 K
Temperature change rate cooling: 6 K/min
Temperature change rate heating: 6 K/min
Test chamber dimensions:
Width: 800 mm, Depth: 800 mm, Height: 950 mm
External dimensions:
Width: 1000 mm, Depth: 1750 mm, Height: 2005 mm
Viewing window:
Width: 300 mm x Height: 540 mm
Test chamber volume: 600 liters
Sound level at 1m distance (free-field measurement): < 69 dB(A)
Left side port: 50 mm diameter
Right side port: 2 x 125 mm diameter
Rated voltage: 3/N/PE AC, 400 V ±10%, 50 Hz
Rated power: 16 kW, Rated current: 30 A
Circuit breaker, customer-provided: 29 A slow-blow
Chamber is water-cooled
Refrigerants: R-404A and R 23
Hnsdpfx Abjxw Uqpedsw
Weight approx. 675 kg
For your safety as the buyer, please note the following information!
The following points are completed on our offered chambers in advance:
1. Functional check and replacement of necessary components
2. Refilling with legal, compliant refrigerants
3. After successful inspection, the chambers are subjected to a test run monitored with calibrated instruments.
Condition: used
Scope of delivery: (see picture)
(Technical data subject to change and errors excepted!)
We would be happy to answer any further questions by phone.
Iklan ini diterjemahkan secara automatik. Mungkin terdapat kesilapan terjemahan.
Type: VTS 7060-5
Halogen Bulb 24V, 50W
Vötsch temperature test chamber with color touch panel and S!MPAT
Temperature range: -70°C to +180°C, water-cooled
Temperature deviation over time: ±0.1 to ±0.5 K
Temperature deviation spatially: ±0.5 to ±2.0 K
Temperature change rate cooling: 6 K/min
Temperature change rate heating: 6 K/min
Test chamber dimensions:
Width: 800 mm, Depth: 800 mm, Height: 950 mm
External dimensions:
Width: 1000 mm, Depth: 1750 mm, Height: 2005 mm
Viewing window:
Width: 300 mm x Height: 540 mm
Test chamber volume: 600 liters
Sound level at 1m distance (free-field measurement): < 69 dB(A)
Left side port: 50 mm diameter
Right side port: 2 x 125 mm diameter
Rated voltage: 3/N/PE AC, 400 V ±10%, 50 Hz
Rated power: 16 kW, Rated current: 30 A
Circuit breaker, customer-provided: 29 A slow-blow
Chamber is water-cooled
Refrigerants: R-404A and R 23
Hnsdpfx Abjxw Uqpedsw
Weight approx. 675 kg
For your safety as the buyer, please note the following information!
The following points are completed on our offered chambers in advance:
1. Functional check and replacement of necessary components
2. Refilling with legal, compliant refrigerants
3. After successful inspection, the chambers are subjected to a test run monitored with calibrated instruments.
Condition: used
Scope of delivery: (see picture)
(Technical data subject to change and errors excepted!)
We would be happy to answer any further questions by phone.